Mass spectrometer unit
Mass number range
1.5 m/z to 1024 m/z
Resolution
R>2M
Scan rate
10,000u/s min.
SIM
64 channels x 64 groups (max.)
Filaments
dual filaments
Evacuation system
Differential pumps
Ion source
turbo-molecular pump with 260 L/s pumping rate
Quadrupoles
turbo-molecular pump with 65 L/s pumping rate
Analysis detector unit
Mass filter
High-precision quadrupoles with pre-rods
Gas chromatograph
GC unit
Shimadzu GC-2010
Data processing unit (minimum specifications)
CPU
Pentium III (600MHz) or higher
Main memory
128MB or larger
OS
Windows 2000/XP
Hard disk
10 GB or more
LAN
Ethernet interface
(100BASE-TX/10BASE-T)
Display
LCD (15 inch) or CRT(17 inch)
resolution 1024x~768 or higher
Printer
Laser printer
Software
Measurement modes
Scan
SIM 64 channels x 64 groups
GC detectors (FID, ECD, FPD, FTD, TCD)
Data analysis
Qualitative processing, quantitative processing
Libraries
NIST, WILEY, DRUG, VOC, Pesticides(EI/NCI)
Calibration curves
1000 compounds max., 64 levels
Grouping
Data conversion
Andi (AIA) conversion, ASCII conversion of quantitation results
Quality control
Pass/fail evaluation of recovery,
degradation check, spectral pattern check, etc.
Security
User designation by username and password,
authority restrictions
Audit trail
Store change histories of method files,
data files, and configuration setting files
Setting conditions
Power requirements
GC single phase
115 VAC, 1800 VA
230 VAC, 2600 VA (high power oven)
MS single phase
100 VAC~240 VAC, 1000 VA
Installed environment
Temperature
18 to 28°C, constant temperature
Humidity
40 to 70% RH (no condensation)
Gas
He (helium), 99.995% min.