Specifications |
|
- Light source: Deuterium lamp
- Wavelength range:
- 190 ~ 600 nm - Cut filter for second order diffracting light is automatically inserted in the 371- 600nm range - Spectral bandwidth: 8 nm
- Wavelength accuracy: ± 1 nm
- Wavelength reproducibility: ± 0.1 nm
- Cell path length: 10nm
- Cell volume: 8ul
- Cell pressure tolerance: 12MPa
- Dual wavelength mode:
- Measuring wavelength: Aritrary two wavelengths in 190- 370 nm or 371-600 nm - Sampling period: 1.2 sec for one wavelength - Response: Selectable in 9 steps corresponding to time constant 0.05, 0.1, 0.5, 1.0, 1.5, 3.0, 6.0, 8.0, 10.0 seconds
- Range: Can be set between 0.0001 and 2.56 AUFS in 0.0001 AUFS steps
- Zero adjustment: Auto zero function, baseline shift function
- Wavelength scanning function: Performs wavelength scanning with flow stopped. Data stored in three files, one of which stores background. Background is subtracted from sample scans. File data is not saved up (lost when power is turned OFF).
- Time Program: Available in detector or by system controller
- Set items: Wavelength ( including dual wavelengths), auto zero, range, marker, response, wavelength scanning, event, polarity, lamp ON/OFF, loop - Number of steps: Maximum 32 steps - Output
- Output for Recorder: 1 mv, 10 mV, maximum output voltage is approx. 20 times as much as full scale - Output for Integrator: 0.5, 1, 2, 4, AU/V in 4 steps
|
|