Instrument Database:
Thermo Scientific - Finnigan TRITON
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Thermo Scientific
|
Categories |
Mass spectrometer: TIMS |
Finnigan TRITON The Finnigan TRITON gives the ultimate precision for isotope ratio measurements on solid samples. It uses a proven thermal ionization source and has a unique variable multicollector platform which can be configured with Faraday detectors and/or miniaturized ion counting detectors for smallest sample sizes. Typical applications are dating of geological samples as well as control of isotopic compositions of nuclear materials. New Technology
The Finnigan TRITON is a completely new thermal ionization mass spectrometer (TIMS) delivering the most precise and accurate isotope ratios ever archieved with TIMS for positive and negative ions.
Through its innovative technology, including Virtual Amplifiers, Dynamic Zoom and all-carbon plug-in Faraday cups, the TRITON sets a new standard for TIMS. Major applications are in isotope geology, geochemistry and planetary research.
The TRITON is the first system built on Thermo Scientific's new multicollector platform. The second member of the product family is the Finnigan NEPTUNE, a multicollector ICP-MS.
- Thermal ionization isotope ratio MS (TIMS) with multicollector
- The most precise and accurate TIMS ever
- Novel patented technology to overcome traditional shortcomings of multicollector MS
- Guaranteed 5 ppm external precision on Sr and Nd
- High abundance filter (RPQ) and multi-ion-counting options
|
Specifications |
|
Mass Resolution (10% valley): >450
Mass Range: 3-310
Acceleration voltage: 10 kV for positive and negative ions
Abundance sensitivity (238U +- 1 u): < 2 ppm in the focal plane of the mass analyzer, < 20 ppb on RPQ channel
Simultaneous collector mass range: 17 %, i.e. 6Li and 7Li
Maximum Collector configuration: 9 Faraday cups + 8 miniaturized ion counters
Amplifier noise: < 0.2 fA rms @ 4s integration time
Inter-channel calibration stability: < 10 ppm within 24 h, < 1 ppm with virtual amplifier mode
Basic system specification: 5 ppm external reproducibility for Nd and Sr
|
|