Instrument Database:
Beijing Rayleigh Analytical Instrument Corporation (BRAIC) - WLD-2C ICP SPECTROMETER
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Beijing Rayleigh Analytical Instrument Corporation (BRAIC)
|
Categories |
Spectrometer ( Atom. ): AES: ICP-AES |
Usage: The ICP Emission Spectrometer is an automated analytical instrument for analyzing microelements and trace elements. It has a wide range of applications, and can be used in the fields of environmental monitoring, chemicals, geology, agriculture, pharmaceuticals, biology, foodstuffs, metallurgy and many others. Features:
- High sensitivity, low detection limit of 10-8~10-10 g/ml.
- High precision with RSD less than 1.5%.
- Minimal matrix effect.
- Wide analyzable concentration range of 4~5 orders of magnitude.
- Simultaneous analysis of multi-elements
- Dynamic background correction to eliminate spectral interference.
- Integrated data collection control board is located inside the computer, ensuring strong resistance to interference.
|
Specifications |
|
RF generator Oscillating frequency 40MHZ Power output 0.7~1.2KW Power output stability Less than 0.5%
Polychromator Paschen-Runge mounting Concave grating: Curvature radius: 750mm Ruled density: 2400/mm Size: 30 x 50 mm Dispersion: 0.55nm/mm (1st order) Wavelength range: 190~500nm Constant temperature: 30°C±0.5°C
Measuring System Measuring mode: sectional integration Measuring accuracy: 0.2% Dynamic range: 106
Computer system PII IBM compatible computer, printer and data collection control board. Application software: Windows 95/98 operating system, standard Windows application software.
Basic functions of the analytical software: Spectrum scanning (manual and automatic); Working curve establishment; Working curve calibration; Background correction ( blank method, dynamic method ); Quantitative analysis ( curve method, standard addition method); Spectral interference correction (coefficient method, kalman filter method).
|
|